Next: T1,T2 classification
Up: Application: tissue classification in
Previous: Application: tissue classification in
Protons in a magnetic field have a microscopic magnetization, and
act like toy tops that wobble as they spin. The rate of the
wobbling, or precession, is the resonance or Larmor frequency. In
the magnetic field of an MRI scanner, there is approximately the
same number of proton nuclei aligned with the main magnetic field
B0 as counter aligned. Still, the aligned position is slightly
favored because the nucleus has a lower energy in this position.
This results in a net macroscopic magnetization pointing in the
direction of B0.
The exposure of the nuclei to a RF radiation (the B1 field) at
the Larmor frequency causes the nuclei in the lower energy state
to jump to the higher energy state. At the macroscopic level, this
causes the net magnetization to spiral away from the B0 field.
In a field of reference rotating with the field, the net
magnetization vector rotates from the longitudinal direction to a
flip angle proportional to the duration of the RF pulse. After
some time, the magnetization vector becomes perpendicular to the
main field B0. In this position, it can be detected by the MRI
scanner. For angles different from 90o, the perpendicular
component of the magnetization vector can still be detected, but
the signal is of course weaker.
MR Imaging is based on the observation of the relaxation that
takes place after the RF pulse has stopped. The return of the
excited nuclei from the high energy to the low energy state is
associated with the loss of energy to the surrounding nuclei.
Macroscopically, this spin-lattice or T1 relaxation is
characterized by the longitudinal return of the net magnetization
to its maximum length in the direction of the magnetic field. This
return is an exponential process of the form of
1 - e-t/T1
(Figure
). The T1 relaxation time is the time
constant of this exponential, i.e. the time needed for the
magnetization to return to
of its original value.
Figure 11.1:
T1 and T2 relaxation
 |
Microscopically, T2 relaxation, or spin-spin relaxation,
occurs when the spins in the high and low energy state exchange
energy but do not loose energy to the surrounding lattice.
Macroscopically, this results in a loss of transverse
magnetization. Once again, T2 relaxation is a exponential process,
in the form of e-t/T2 (Figure
), and the T2
time is the time needed for the transverse magnetization to lose
of its value. In pure water, the T2 and T1 times are
approximately identical. For biological material, the T2 time is
considerably shorter than the T1 time.
Figure 11.2:
T1 and T2 MR images.
 |
By varying imaging parameters such as TR (repetition time) and TE
(echo time), it is possible to weight the IRM signal to produce
T1-, T2- or PD-weighted (proton density) images. From a medical
perspective, it means that MR Imaging can provide multiple
channels to observe the same anatomy. For instance, Figure
shows T1- and T2-weighted images of the same brain.
Different tissues appear differently in both images. White matter
appears in a light grey in T1 and a dark grey in T2. Grey matter
appears grey in both images. The Cerebro-Spinal Fluid (CSF)
appears black in T1 and white in T2. The background of the image
(air) appears black in both images.
Next: T1,T2 classification
Up: Application: tissue classification in
Previous: Application: tissue classification in
Olivier Cuisenaire
1999-10-05